Our Process Optical Dissolved Oxygen Electrode enables optical dissolved oxygen measurement technology on existing 4-20mA or 0 to 5VDC test systems. You can now buy just the electrode and not have to invest in a whole new analysis system.
This technology is based on the proven method whereas oxygen concentration is determined by the luminescent quenching of a dye embedded on the sensor surface. Unlike polarographic or galvanic sensors, this optical sensor does not require filling solutions nor a “warm-up” time.
Features:
* Affordable, complete electrode system
* 4-20mA output for existing controllers/monitors
* No electrolyte solution to replenish
* Replaceable membranes
* No calibration or replacement membranes for at least 1 year in most applications
* Saltwater compensation with toggle switch
* Sample flow
Range…………………. Range: 0-20 ppm DO (0-200% saturation)
Accuracy……............. ±0.1 ppm below 1 ppm
±0.2 ppm above 1 ppm
Response Time……... T90 <30 seconds
Resolution…………… 0.1 ppm
Operating Temperature…………. 0-65 ºC
Materials
Body Housing…PVC/Stainless Steel 316
Membrane…PET
Guard…SS316 (options exist)
Dimensions…………..
225mm (8.9-inch) overall length
70.5mm (2.78-inch) insertion length
42.2mm (1.66-inch) body OD
27.9mm (1.1-inch) insertion OD
1” NPT front and back fittings
Input Power……….. 5-16V
Output Signal……….. 4-20mA or 0-5V